Analytical and Preliminary Clinical Evaluation of a Point-of-Care Device for Urinary Albumin Measurement
Wanna Chaijaroenkul
Graduate Program in Bioclinical Sciences, Chulabhorn International College of Medicine, Thammasat University, Pathum Thani 12120, Thailand.
Napaporn Youngvises
Drug Discovery and Development Center, Office of Advanced Science and Technology, Thammasat University, Pathum Thani 12120, Thailand.
Artitaya Thiengsusuk
Drug Discovery and Development Center, Office of Advanced Science and Technology, Thammasat University, Pathum Thani 12120, Thailand.
Tullayakorn Plengsuriyakarn
Graduate Program in Bioclinical Sciences, Chulabhorn International College of Medicine, Thammasat University, Pathum Thani 12120, Thailand.
Wanchai Meesiri
Bangkok High Lab Co., LTD., Bang Khen District, Bangkok 10220, Thailand.
Kesara Na-Bangchang *
Graduate Program in Bioclinical Sciences, Chulabhorn International College of Medicine, Thammasat University, Pathum Thani 12120, Thailand and Drug Discovery and Development Center, Office of Advanced Science and Technology, Thammasat University, Pathum Thani 12120, Thailand.
*Author to whom correspondence should be addressed.
Abstract
Microalbuminuria is an important marker for early assessment of chronic kidney disease and related renal complications. This study describes the development and preliminary validation of MyμAlbumin, a connected point-of-care device designed to measure urinary albumin using an immunoturbidimetric assay. The system uses a 340 nm ultraviolet light-emitting diode, photodiode detection, microcontroller-based signal processing, and a disposable microcuvette requiring a total reaction volume of 125 μL, including 10 μL of urine sample or albumin standard. Albumin-specific immune complex formation increases turbidity, and the resulting absorbance signal is converted into albumin concentration using a polynomial calibration model. Analytical evaluation was performed using artificial urine samples spiked with albumin across the calibration range and at selected concentrations of 20, 100, and 300 mg/L. The reported calibration performance showed strong polynomial correlations for both the MyμAlbumin device and the laboratory spectrophotometer. In artificial urine, MyμAlbumin showed intra-day and inter-day precision values ranging from 0.08% to 1.97% coefficient of variation, with percentage deviation from the mean ranging from −4.82% to 4.28%. The limit of quantification was 5 mg/L. Clinical applicability was assessed using urine samples from 25 patients and compared with laboratory spectrophotometric and hospital-based immunoturbidimetric methods. Median percentage deviations were 1.45% and 1.58% against the two reference methods, respectively, although wider individual ranges were observed. These findings indicate that MyμAlbumin may provide a compact and rapid approach for urinary albumin assessment, but larger clinical validation studies are required.
Keywords: Microalbuminuria, urinary albumin, point-of-care testing, immunoturbidimetry, chronic kidney disease, microcuvette, albumin quantification, artificial urine, clinical validation, connected diagnostics